| CLD | NDIR / NDUV / MPA |
| NOX | ● / ● |
| SO2 | ● / ● |
| CO | ● |
| CO2 | ● |
| O2 | ● |
| NH3 | ● |
| Component | Principle / Range |
| NOX (Low) | CLD / 0-5/5000 ppm (dual range) |
| NOX | NDIR / 0-200/5000 ppm (dual range) |
| SO2 (Low) | NDUV / 0-10/1000 or 0-5/1000 ppm |
| SO2 | NDIR / 0-200/5000 ppm (dual range) |
| CO (Low) | NDIR / 0-5/500 ppm (dual range) |
| CO | NDIR / 0-100/5000 ppm (dual range) |
| CO (High) | NDIR / 0-5 vol% |
| CO2 | NDIR / 0-5/50 vol% |
| O2 | MPA / 0-5/25 vol% |
| NH3 | CLD / 0-10/100 ppm (dual range) |
| Repeatability | ±0.5% full-scale ±0.1% full-scale (optional range) |
| Zero drift | ±1.0% full-scale per week ±2.0% full-scale per week (optional range) |
| Span drift | ±2.0% full-scale per week |
| Linearity | ±1.0% full-scale per week ±2.0% full-scale per week (optional range) |
| Communications | RS-232C, RS-485 |
| Serial Ports | USB, Ethernet |
| Ambient temperature | Standard: +5°C to 40°C, Option: -5°C to 50°C |
| Response time | 45 seconds max. at system inlet (T90) |
| Total interference | ±2.0% full-scale |
| Dimensions [system in cabinet] | [inches] 25.6 (W) x 32 (D) x 75 (H) [mm] 651 (W) x 813 (D) x 1905(H) |
| Weight [system in cabinet] | Approx. 700 lb (318 kg) |
|---|