HE-960LF / FS-09F-1/2
规格参数。Product name Conductivity meter Model HE-960LF Measurement method Electrode type(2-electrode method) Temperature specifications sensor Platinum resistance 1000Ω/℃ Measu
广西科米瑞实验仪器设备有限公司生产销售 HORIBA HE-960LF / FS-09F-1/2,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表
HE-960LF / FS-09F-1/2
| Product name | Conductivity meter |
|---|---|
| Model | HE-960LF |
| Measurement method | Electrode type(2-electrode method) |
| Temperature specifications sensor | Platinum resistance 1000Ω/℃ |
| Measurement range | Cell constant / 0.1/cm / 1.0/cm |
| Conductivity | 0.00 to 20.00μS/cm / 0.0 to 200.0μS/cm |
| 0.0 to 200.0μS/cm | 0 to 2000μS/cm |
| 0 to 1000μS/cm | 0 to 10000μS/cm |
| Temperature | 0 to 100℃ Actual temperature measuring range depend on the performance of the sensor connected. |
| Concentration conversion | Option 1,2 / 0 to 10.000% Conversion formula is defined by user (temperature compensation and concentration conversion) |
| Repeatability | Cell constant / 0.1/cm / 1.0/cm |
| Conductivity | 0.00 to 20.00μS/cm range / 0.0 to 200.0μS/cm range / Within ±0.5% full scale |
| 0.0 to 200.0μS/cm range | 0 to 2000μS/cm range |
| 0 to 1000μS/cm range | 0 to 10000μS/cm range / Within ±1.0% full scale |
| Temperature | ±0.5℃ |
| Condition | Equivalent input |
| Linearity | Cell constant / 0.1/cm / 1.0/cm |
| Conductivity | 0.00 to 20.00μS/cm range / 0.0 to 200.0μS/cm range / Within ±0.5% full scale |
| 0.0 to 200.0μS/cm range | 0 to 2000μS/cm range |
| 0 to 1000μS/cm range | 0 to 10000μS/cm range / Within ±1.0% full scale |
| Temperature | ±0.5℃ |
| Condition | Equivalent input |
| Transmission output | Number of output:4 DC4 to 20mA / 0 to 20mA input/output isolated type Maximum load resistance: 900Ω Transmission output range: Free setting within measuring range (Negative terminals of each transmission output channel are connected inside and thus have the same electrical potential.) |
| Contact output | Number of relay:5 ALARM contact R1,R2,R3 and R4 Contact type relay: Relay contact, SPST (1a) Contact rating:240 V AC,1 A or 30 V DC, 1 A(resistance load) Contact function Upper or lower ON/OFF alarm each measurement items. Contact action Closed when status is in the event. Opened when any erroneous status is normal or power is down. R1,R2 and R3 share a common terminal. Self diagnosis contact RF Contact type: relay contact , SPDT (1c) Contact rating:240 V AC,1 A or 30 V DC, 1 A(resistance load |
| Contact input | Number of input:1 |
| Communication output | RS-485 communication |
| Self-check | Sensor diagnosis (Short-circuit and disconnection of the temperature sensor). Converter error |
| Temperature compensation of conductivity | ● Arbitrary temperature coefficient entry (reference temperature: 25℃, temperature coefficient: 0%/℃ to 5%/℃) ● Arbitrary temperature compensation formula entry (reference temperature: 25℃, formula is defined by user) ● No temperature compensation is performed. |
| Temperature compensation range | 0 to 100℃ The temperature compensation under 0℃ and over 100℃ is expanded by a function |
| Ambient temperature | 5 to 45℃ |
请填写必填信息,我们将在工作日尽快回复。