VS70-PDA-HDR
规格参数。Spectral Coverage UV-VIS (200–800 nm) with 477 l/mm grating, VIS (280--700 nm) with 582 l/mm grating, UV-NIR (200–985 nm) with double blazed 365 l/mm grating. On reques
广西科米瑞实验仪器设备有限公司生产销售 HORIBA VS70-PDA-HDR,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表
VS70-PDA-HDR
| Spectral Coverage | UV-VIS (200–800 nm) with 477 l/mm grating, VIS (280--700 nm) with 582 l/mm grating, UV-NIR (200–985 nm) with double blazed 365 l/mm grating. On request, other spectral ranges available; different mechanical integration might be needed |
|---|---|
| Spectrometer Input | Fiber-coupled: FC or SMA; Free space: 12-25-37-50-62-75-100-125-150-200 μm slits; Other options available upon request |
| Spectral Resolution | 2 nm for 365 l/mm grating, 1 nm for 477 l/mm grating, 0.8 nm for 582 l/mm grating (25 μm slit for all) |
| Average Spectral Dispersion | 30.9 nm/mm for 365 l/mm grating, 23.7 nm/mm for 477 l/mm grating 16.3 nm/mm for 582 l/mm grating |
| Average Digital Resolution with 25 (50) μm pixel width | 0.8 (1.5) nm/px for 365 l/mm grating, 0.59 (1.2) nm/px for 477 l/mm grating, 0.4 (0.8) nm/px for 582 l/mm grating |
| Focal Length | 70 mm |
| Options | Selection of high grade sensors: Hamamatsu® S1590x and S1012x families (other PDA sensors available upon request); Input port: SMA, FC, free space, custom input (e.g., round-to-line fiber bundle) |
| F/# | F/2.8 |
| Stray Light Rejection Typical | 0.01% for UV-VIS configuration measured at 400 nm and 700 nm with a 550 nm (40 nm band) bandpass filter, 31 μm slit-width |
| Wavelength Accuracy | 0.2 nm |
| Software** | LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration) |
| Detector Model | Hamamatsu S10124-1024 o r S15909-1024 with high UV sensitivity. |
| Sensor Format | 1024 x 1 pixels |
| PDA Pixel Size | 25 μm x 2500 μm |
| PDA Active Area | 25.6 mm x 2.5 mm |
| PDA Spectral Response | S15909-1024 has improved spectral response (NO FRINGES) compared to S10124-1024. (See figure below) |
| Sensor Temperature | Uncooled |
| Full Well Capacity | 624 Me- for S15909-1024 and 468 Me- for S10124-1024 |
| Readout Noise | 14,700 e- (typical) for S15909-1024. S10124-1024 not yet evaluated |
| Digitization | 18-bit |
| Dynamic Range (FW/RN) | 42,500:1 (typical) for S15909-1024. S10124-1024 not yet evaluated |
| Max SNR (√FW) | 24,980:1 (typical) for S15909-1024 and 21,600:1 (typical) for S10124-1024 |
| Non-linearity (measured on each system) | < 0.4% (corrected) |
| Dark Current (at 25C ambient) | 190,000 e-/pix/s (typical) for S15909-1024 and 625,000 e-/pixel/s (typical) for S10124-1024 |
| Communication | USB 2.0 High-speed, (USB 3.0 customizable) |
| Environmental Conditions | Operating temperature 15° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
| Power Requirements | 5 VDC @ 0.13 A, USB-powered or Molex connector |
| Detector Model | Hamamatsu S15908-512 or S10121-512 with high UV sensitivity. |
| Sensor Format | 512 x 1 pixels |
| PDA Pixel Size | 50 μm x 2500 μm |
请填写必填信息,我们将在工作日尽快回复。