产品概述与应用
Waterproof and Dustproof Housing - IP67 rated, ideal for rugged harsh conditions. Plus it floats!
Multi-Point Push-button Calibration - up to 5 points with Auto-Buffer Recognition for quick, error-free calibration
Choice of pH Buffer Sets - USA, NIST or custom buffer standards for wider applications
User-Customization - for enhanced flexibility and viewing of electrode diagnosis - pH slope/offset/mV base value, previous calibration points
Built-in Real-time Clock - stamps stored data and calibration data with time & date - meets standards for GLP purposes
Expanded Memory - stores and recalls up to 50 readings with temperature
Automatic Temperature Compensation (ATC) - ensures highest accuracy under fluctuating temperatures
Stability Indicator - prompts whenever reading stabilizes
Auto-Hold Function - automatically freezes reading when stabilized for easy viewing and recording
Auto-Power Off - conserves battery power and saves money
Self-Diagnostic - includes message codes for easy troubleshooting
Large Custom LCD Display - provides optimum viewing with simultaneous readings and user-friendly icons
Optional Rugged Carrying Case - enables convenient on-site meter calibration and measurement
Industrial: Ideal for checks in water conditioning plants, cooling towers, plating and finishing operations, food processing (for HACCP compliance), water testing, printing and chemical industries, rinse water analysis, water/wastewater treatment and chemical process verification. Use with ORP electrode to monitor plating operations, and other redox measurements.
技术参数
| 型号/系列 | CyberScan pH 310 |
|---|---|
| pH Range | -2.00 to 16.00 pH |
| Resolution & Accuracy | 0.01 pH ±0.01 pH |
| mV Range | ±199.9 mV (extended); ±1999 mV |
| Resolution & Accuracy | 0.1 mV & ±0.2 mV (within 199.9 mV); 1 mV & ±2 mV (beyond ±200 mV) |
| Temperature Range | 0 to 100.0 °C (32.0 to 212.0 °F) |
| Resolution & Accuracy | 0.1 °C/°F ±0.3 °C/±0.5 °F |
| No. of Calibration Points | 1 to 6 |
| pH Buffer Options | USA: 1.68, 4.01, 7.00, 10.01, 12.45; NIST: 1.68, 4.01, 6.86, 9.18, 12.45; DIN: 1.09, 3.06, 4.65, 6.79, 9.23, 12.75 |
| Memory | 50 data sets |
| Temperature Compensation | Automatic / Manual (0 to 100 °C) |
| Date/Time Stamping | Yes |
| Special Functions | Auto-Hold; Auto Power-Off after 30 minutes; Stability Indicator; Self-diagnostics messages |
| Power | 4 ‘AAA’ x 1.4V batteries |
咨询仪器选型与报价
提供产品名称、型号、数量和使用要求,我们将协助确认参数与配置。