广西科米瑞实验仪器设备有限公司应用与技术支持
返回上一页
Thermo Scientific Alpha COND 2000 Thermo Scientific Alpha COND 2000
产品与应用图片。
实验室仪器

Thermo Scientific Alpha COND 2000

Eutech Instruments

是 Eutech Instruments 电导率测量仪,原始说明为:Measure from ultra-pure water to highly conductive samples with the Alpha COND 2000 controller/transmitter: this conductivity controller accepts 4-cell and 2-cell electrodes, and features programmable temperature coefficients to give you precise temperature compensation.

电话咨询这款仪器

微信扫码咨询

请发送仪器型号与应用需求,咨询选型及报价。

微信咨询二维码 13422079856
查看技术参数

根据您的样品与应用需求,咨询适用配置。

测量方法

产品概述与应用

Measure from ultra-pure water to highly conductive samples with the Alpha COND 2000 controller/transmitter: this conductivity controller accepts 4-cell and 2-cell electrodes, and features programmable temperature coefficients to give you precise temperature compensation.

Wide Measurement Range - Measures seven ranges, from 0 to 2 µS/cm to 0 to 1000 mS/cm

High Resolution & Accuracy - Readings of up to three-decimal resolution at ±1 % accuracy

Advanced Temperature Compensation - Adjustable temperature coefficients with programmable linear compensation (0.0 % to 10.0 %) and pure water compensation

Electrode Diagnostics - Meter displays electrode status after each calibration

Flexible Cell Constants - User-adjustable cell constant values (k = 0.005 to k = 9.999)

Superior Electrode Design - 4-cell electrode prevents polarizing effects and electrode fouling in samples with high conductivity

Versatile Temperature Sensors - Accepts 2- or 3-wire, Pt100 or Pt1000 RTD sensors for automatic temperature compensation

Independent Calibration - Allows user to input cell constant corresponding to connecting electrode independently during calibration

Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator.

Use in applications involving agriculture, chemical processing, boiler and water heaters, wafer-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries, and bleach manufacturing.

Use to treat batches of incoming process water, ultrapure water, boiler and feed water control.

Hook to recorder to document data for regulatory compliance.

技术概览

技术参数

型号/系列Thermo Scientific Alpha COND 2000
Range… to 2.000 µS/cm ; … to 20.00 µS/cm ; … to 200.0 µS/cm ; … to 2000 µS/cm ; … to 20.00 mS/cm ; … to 200.0 mS/cm ; … to 1000 mS/cm
Resolution0.001 µS/cm ; 0.01 µS/cm ; 0.1 µS/cm ; 1 µS/cm ; 0.01 mS/cm ; 0.1 mS/cm ; 1 mS/cm
Accuracy±1 % of full scale reading (±2 % >500 mS/cm)
Range-10.0 to 125.0 °C (14.0 to 257.0 °F)
Resolution0.1 °C / 0.1 °F
Accuracy±0.5 °C / ±1.0 °F
SensorPt100/Pt1000 (jumper selectable); 2 or 3 wire
CompensationAuto/manual (normalized at 25 °C)
Set Point 1 (SP1) / Set Point 2 (SP2)… to 2.000 µS/cm or … to 20.00 µS/cm or … to 200.0 µS/cm or … to 2000 µS/cm or … to 20.00 mS/cm or … to 200.0 mS/cm or … to 1000 mS/cm
Switching Conductivity Hysteresis0 to 10 % of full scale
Function (Switchable)P/PI control (pulse length/pulse frequency/proportional integral); limit control; off
Adjustable Period With Pulse Length Controller0.5 to 20 sec
Adjustable Period With Pulse Frequency Controller60 to 120 pulse/min
Integral Action Time (IAT)0 to 999.9 min
Pickup/Dropout Delay0 to 2000 sec
Wash Cycle0.1 to 200.0 hr
Wash Duration1 to 2000 sec
Contact Outputs1 SPDT, 3 SPST relays
Switching Voltage/Current/Power1/8 HP: at 125 VAC / max. 0.74 A / max. 93 VA 1/8 HP: at 250 VAC / max. 0.37 A / max. 93 VA
Function (Switchable)Steady or fleet (pulse)
Pickup Delay0 to 2000 sec
Switching Voltage/Current/Power1/8 HP: at 125 VAC / max. 0.74 A / max. 93 VA 1/8 HP: at 250 VAC / max. 0.37 A / max. 93 VA
Transmitter FunctionTwo 0/4 to 20 mA scalable outputs for Conductivity and temperature, galvanically isolated
CU 22 Function22 mA current output
应用咨询

咨询仪器选型与报价

提供产品名称、型号、数量和使用要求,我们将协助确认参数与配置。

联系科米瑞
Kemirui / 应用咨询

一起讨论您的测量需求。

您的信息用于处理本次咨询。