HC-200N HC-200N
产品与应用图片。
实验室仪器

HC-200N

HORIBA

Reduces power consumption and costs at sewage treatment plants and other water treatment facilities. The proprietary technology enables high durability and long-term stability.

电话咨询这款仪器

微信扫码咨询

请发送仪器型号与应用需求,咨询选型及报价。

微信咨询二维码 13422079856
查看技术参数

根据您的样品与应用需求,咨询适用配置。

测量方法

产品概述与应用

产品说明

Reduces power consumption and costs at sewage treatment plants and other water treatment facilities. The proprietary technology enables high durability and long-term stability.

规格参数

Product name Ammonia/Nitrate Nitrogen Meter
Converter model HC-200N
Sensor model AM-2000-A: Ammonia nitrogen sensor
Sensor tip model 7691: Ammonium ion tip, 7692: Potassium ion tip for compensation, 7211: Reference tip
Measurement range Concentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L) Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolution Concentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L Temperature: 0.1℃
Repeatability Concentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃) Temperature: ±0.3℃(equivalent input)
Temperature compensation Compatible temperature element: platinum resistor: 1 kΩ (0℃) Temperature measurement range: 0℃ to 40℃ Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Potassium ion concentration compensation Compensation range: the potassium ion concentration is not more than 10 times the ammonium ion concentration and 1000 ppm or less. Compensation error: ±20%(reading)
Measurable pH range pH 4.0 to 8.5
Calibration Reference solution calibration of the ammonia nitrogen (2points) Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L. Additional functions: automatic defect calibration judgment, calibration history
Adjustment Sample adjustment using hand analysis (1 point, 2 points*1), standard curve input function (primary system) Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detection Calibration error, sensor diagnosis error, converter error
Sensor model NO-2000: Nitrate nitrogen sensor
Sensor tip model 7693: Nitrate ion tip, 7694: Chloride ion tip for compensation, 7211: Reference tip
Measurement range Concentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L) Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolution Concentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L Temperature: 0.1℃
Repeatability Concentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃) Temperature: ±0.3℃(equivalent input)
Temperature compensation Compatible temperature element: platinum resistor: 1 kΩ (0℃) Temperature measurement range: 0℃ to 40℃ Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Chloride ion concentration compensation Compensation range: when the chloride ion concentration is 100 times greater than the nitrate ion concentration, a correction factor must be applied to account for interference. Compensation error: ±20%(reading)
Measurable pH range pH 2.0 to 12.0 When the pH is between pH 2.0 to pH 4.0 or pH 9.0 to pH 12.0, use sensor cap AM-CAP-PVC. Chloride ion tip can be used in the pH 2.0 to pH 8.0. However, when used at pH 2.0 to pH 4.0, the lifetime of the tip may be shortened.
Measurement inhibitors Component: if the surfactant concentration is 10mg/L or higher, sensor performance may degrade, and measurement may not be possible.
Calibration Reference solution calibration of the ammonia nitrogen (2points) Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L. Additional functions: automatic defect calibration judgment, calibration history
Adjustment Sample adjustment using hand analysis (1 point, 2 points), standard curve input function (primary system) Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detection Calibration error, sensor diagnosis error, converter error
Transmission output 3 points, 4 mA to 20 mA DC, input/output isolation type, 900Ω(max) Select 3 items from below. Ammonia nitrogen concentration: arbitrary setting within the measurement range is possible. Nitrate nitrogen concentration: arbitrary setting within the measurement range is possible. Chloride ion concentration: arbitrary setting within the measurement range is possible. Temperature reading of the ammonia nitrogen meter: arbitrary setting within the range from ⁻10℃ to 110℃ is possible. Temperature readi
Contact output 5 points, No-voltage contact output, relay contact, SPST(1a),SPDT(1c) R1,R2: relay contact SPST(1a) Contact function: selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, and cleaning output (closed at alarm operation, open under normal conditions and in a power off state). FAIL: relay contact SPDT(1c) Contact function: error alarms for values outside the measurement range and self check output can be set. Delay time: 0 seconds to 600 seconds R3,R4: re
Cleaning output 2 points, Voltage contact output (connection power supply voltage output) Relay contact, SPDT(1c), SPST(1a)
Contact input 1 point, Contact type: open collector no-voltage “a” contact
Communication function Type: RS-485 Signal form: Two wire type, the power source of the communication is isolated from the power of measurement circuit.(The power source of transmission and communication are not isolated.)
技术概览

技术参数

品牌HORIBA
型号/名称HC-200N
Product nameAmmonia/Nitrate Nitrogen Meter
Converter modelHC-200N
Sensor modelAM-2000-A: Ammonia nitrogen sensor
Sensor tip model7691: Ammonium ion tip, 7692: Potassium ion tip for compensation, 7211: Reference tip
Measurement rangeConcentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L) Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolutionConcentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L Temperature: 0.1℃
RepeatabilityConcentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃) Temperature: ±0.3℃(equivalent input)
Temperature compensationCompatible temperature element: platinum resistor: 1 kΩ (0℃) Temperature measurement range: 0℃ to 40℃ Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Potassium ion concentration compensationCompensation range: the potassium ion concentration is not more than 10 times the ammonium ion concentration and 1000 ppm or less. Compensation error: ±20%(reading)
Measurable pH rangepH 4.0 to 8.5
CalibrationReference solution calibration of the ammonia nitrogen (2points) Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L. Additional functions: automatic defect calibration judgment, calibration history
AdjustmentSample adjustment using hand analysis (1 point, 2 points*1), standard curve input function (primary system) Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)